japanesechinese


NF Corporation
network contact us site map
Google  
home about nf products support news
Measurement Instruments Power Supply and Power Control produts Function Modules Customized Products
home > Products > Measurement Instruments > LCR Meter > ZM2371/ZM2372 > Features
Product Information

LCR Meter
LCR Meter
ZM2371/ZM2372
mark Basic accuracy: 0.08 %, display resolution of 6 digits (max.)

Measurement speed: Max. 2 ms at 1 kHz
space
Measurement speed is selectable between 5 levels: RAP (rapid) , FAST, MED, SLOW and VSLO (very slow). When set to RAP, high-speed measurement at 2 ms (1kHz) or 10ms (120 Hz) can be performed.

Measurement frequency: 1 mHz to 100 kHz, 5-digit resolution

Measurement signal level: Max. 5 Vrms, 3-digit resolution

Measurements parameters: Lp, Ls, Cp, Cs, Rp, Rs, |Z|, |Y|, G, Q, D, theta, X, B, Rdc

mark Internal DC bias: 0 V to +2.50 V

mark Constant-voltage/constant-current drive (ALC:Auto Level Control)

mark DC resistance measurement

4-terminal contact check function and a handler interface for production lines (ZM2372)
space
ZM2372 is equipped with contact check function to prevent mis-measurement or incorrect sorting due to defective contact between the measurement tip and component. It checks contact with 4 measurement terminals, and if a contact defect occurs, it halts measurement/sorting and determines the contact defect. This function contributes to improved yield.
space
contact check

Step1:
Checks whether there is continuity along the dotted lines (contact check).
Determines there is a defect if contact is defective.

ya
Once the contact failure is detected, ZM2372 allows the measured value “NC” to appear without performing normal measurement.
The additional time necessary for 4-terminal contact check is approx. 4 ms.
ya
Step2: Measures along the dotted lines (main measurement)
 


mark Other functions:
Correction functions (Open, Short, Load and Cable length), Deviation measurement, Comparator function etc.

mark Interfaces
space
  • RS-232
  • USB
  • GPIB
    (Standard for ZM2372, not provided for ZM2371)
  • Handler interface
    (Standard for ZM2372, not provided for ZM2371)

zm2372_rear panel

mark Application software
space

Supports the remote control of ZM2372/ZM2372.
Main functions:
• Sets and acquires measurement parameters
• Controls measurement and acquires measurement result
• Outputs the acquired measurement result to a file


ap_software2
ap_software2
Main screen
Sample graph display using Microsoft Excel




back
top of this page